  The basic principle of the dielectric constant tester is to adopt the high-frequency resonance method, and it provides universal, multi-purpose, and multi-range impedance testing.It uses a single-chip computer to control the instrument, and the measurement core adopts new technologies such as frequency digital locking, automatic setting of standard test frequency points, automatic search of resonance points, automatic conversion of Q value ranges, and digital display of values. It improves the tuning circuit, reducing the residual inductance of the tuning test circuit to a minimum, and retains the automatic amplitude stabilization and other technologies of the original Q meter, making the new instrument more convenient to use and more accurate in measurement. The instrument can measure the Q value of high-frequency inductors or resonant circuits, the inductance and distributed capacitance of inductors, the capacitance and loss tangent of capacitors, the high-frequency dielectric loss of electrical materials, the effective parallel and series resistance of high-frequency circuits, and the characteristic impedance of transmission lines at higher test frequencies.

Technical Specifications


1. Q value measurement

a. Q value measurement range: 5 to 999.

b. Q value range gears: 30, 100, 300, 999, automatic gear shifting.

c. Nominal error

Frequency range: 25kHz to 10MHz;

Intrinsic error: ≤5% ± 2% of full scale; Operating error: ≤7% ± 2% of full scale;

Frequency range: 10MHz to 50MHz;

Intrinsic error: ≤7% ± 2% of full scale; Operating error: ≤10% ± 2% of full scale.

2. Inductance measurement

a. Measurement range: 0.1μH to 1H.

b. Gears: Divided into seven ranges.

0.1 to 1μH, 1 to 10μH, 10 to 100μH,

0.1 to 1mH, 1 to 10mH, 10 to 100mH, 100mH to 1H.

3. Capacitance measurement

a. Measurement range: 1 to 460pF (for capacitance above 460pF, see usage rules);

b. Capacitance adjustment range

Main tuning capacitor: 40 to 500pF;

Accuracy: below 150pF ±1.5pF; above 150pF ±1%;

Fine tuning capacitance: -3pF to 0 to +3pF;

Accuracy: ±0.2pF.

4. Oscillation frequency

a. Oscillation frequency range: 25kHz to 50MHz;

b. Frequency gears:

25 to 74kHz, 74 to 213kHz, 213 to 700kHz, 700kHz to 1.95MHz,

1.95MHz to 5.2MHz, 5.2MHz to 17MHz, 17 to 50MHz.

c. Frequency error: 2×10 ± 1 digit.

5. Q value pass indication preset function, preset range: 5 to 999.

6. Normal working conditions of the instrument

a. Ambient temperature: 0℃ to +40℃;

b. Relative humidity: <80%;

c. Power supply: 220V ± 22V, 50Hz ± 2.5Hz.

7. Sample size

Disc shape: thickness 2+0.5mm, diameter Φ30~40mm (for ε<12), Φ25~35mm (for ε=12~30), Φ15~20mm (for ε>30)

8. Others

a. Power consumption: about 25W;

b. Net weight: about 7kg;

c. Dimensions: (l×b×h) mm: 380×132×280.



Note: The above information and technical parameters are compiled and released by Yangzhou Darui Electric Co., Ltd.
WeChat QR Code
💬
💬 Message ✕

Hello! Have you found what you need? Contact us directly online!

Leave a Message for Consultation
Yangzhou Darui Electric Co., Ltd.
主营产品:Dielectric Loss Tester、Transformer Ratio测试仪、Loop Resistance Tester、高压None线核相仪、Insulation Resistance Tester、Partial Discharge Tester、Cable Fault Tester、DC Resistance Tester等。
☎ Phone: 18952707000(关经理)📞 Landline: 0514-88770377✉ Email: 195214804@qq.com
Verification Code
By submitting, you agree to ourTerms of ServiceandPrivacy Policy
🏠 网站Home📞 一键拨号✉ Contact Us