I. Overview
SXJS-IV is an advanced instrument for measuring dielectric loss (tgδ) and capacitance (Cx), used under power frequency high voltage,to measure the dielectric loss (tgδ) and capacitance (Cx) of various insulating materials, insulating bushings, power cables, capacitors, transformers, and other high-voltage equipment. It replaces the QSI high-voltage bridge, featuring simple operation, Chinese display, printing, easy use, no conversion needed, built-in high voltage, strong anti-interference capability, and short test time. It is the third-generation intelligent dielectric loss tester from our factory.
SXJS-IV Technical Specifications
1. Environment: -5°C to 40°C (avoid prolonged sunlight on the LCD screen)
2. Relative humidity: 30% to 70%
3. Power supply: Voltage: 220V±10%, frequency 50±1Hz
4. Dimensions: Length*Width*Height = 500mm*300mm*400mm
5. Weight: 28kg
6. Output power: 1.5KVA
7. Display resolution: 3 digits, 4 digits (internally all 6 digits)
8. Test methods: normal connection, reverse connection, external test voltage method
9. Measurement range: Internal test voltage: ≤60000PF (for 10KV voltage, greater than 30000PF)
External test voltage: determined by the output power of the external test transformer
10. Basic measurement error:
Dielectric loss (tgδ): 1%±0.04%
Capacitance (Cx): 1%±1pF

Note: The above information and technical parameters are compiled and published by Yangzhou Darui Electric Co., Ltd.